Previous ChapterIndex| Next Page |Next Chapter


Spectophotometric Analysis of Drugs


Theory

Absorption by molecules in solution produces changes in electronic transitions as well as vibrational and rotational changes. For example the carbonyl group bonds contain sigma and pi electrons. These electrons may transition from bonding to antibonding levels.

Sigma and pi bonding and antibonding levels

Figure II-8 Sigma and Pi Bond Levels

Each of these transitions would result in a single peak in the absorbance / wavelength spectrum except for the broadening effect of the rotational and vibrational transisitions.

Absorbance/wavelength spectrum

Figure II-9 Plot of Absorbance versus Wavelength

As light pases through a compound in solution the intensity is reduced.

Light absorbed as it passes through a soultion

Figure II-10 Light Absorbed through a Solution

The longer the pathlength the more light is absorbed. Also, the higher the concentration of compound in solution the more light is absorbed. Absorbance is proportional to pathlength and the concentration (Beer-Lambert's law)

Thus, A = a ¥ b ¥ c where a = absorptivity (epsilon - molar absorptivity includes pathlength and wavelength), b = pathlength (commonly 1 cm), and c = concentration (molar if molar absorptivity)

If b is 1 cm and c is in g/100ml the absorptivity is given as A1%1 cm at wavelength (lambda).

Instrumentation

Single Beam Spectrophotometer

Single Beam Spectrophotometer

Figure II-11 Turner Model 330 Spectrophotometer

The Turner model 330 single beam spectrophotometer has a cell holder for the sample and dials for zero adjustment, 100% transmission and wavelength. Absorbance is read from the upper scale on the meter.

Calibration of the Spectrophotometer

The spectrophotometer should be turned on at least 30 minutes before calibration. Adjust the wavelength to the required setting. With the 'Press for Zero Set' button depressed adjust the 'Zero Adjustment' dial until the meter needle is aligned with 0 % transmission (lower scale). Then place an appropriate blank in a cell in the cell holder and adjust the '100 % Transmission' dial until the meter needle is aligned with 100 % transmission (lower scale). The blank can be replaced with the sample(s) of interest and the absorbance read from the upper scale. This procedure should be made for each wavelength of interest. The spectrophotometer should remain on until all the required readings are made.

Double Beam Spectrophotometer

A Double beam spectrophotometer

Figure II-12 Schematic of a Double Beam Spectrophotometer

Application

Laboratory Exercises

Analysis of Drugs by Visible Spectroscopy

Example Standard Curve at 570 nm

Figure II-13 Absorbance versus Concentration

Extraction of Salicylic Acid

One Compartment - IV Bolus

One Compartment - IV Infusion

One Compartment - Oral

One Compartment Model - IV Bolus - Multiple Dose


This page was last modified: 12 February 2001

Copyright 2001 David W.A. Bourne


Previous ChapterIndex| Next Page |Next Chapter